Diagnostic accuracy of a pH stick, modified to detect gastric lipase, to confirm the correct placement of nasogastric tubes
  Objective The correct placement of a nasogastric feeding tube is usually confirmed by establishing that an aspirate is acidic using a pH stick. However, antacid medication and achlorhydria can cause false negative pH tests that may delay feeding and increase resource use. The purpose of this study was to evaluate a modified pH stick designed to detect gastric lipase and therefore reduce false negative
tests.
Methods In this prospective observational study, a convenience sample of adult patients who had either gastric and oesophageal samples taken during routine diagnostic gastroscopy (n=97) or bronchial and saliva samples taken during a bronchoscopy (n=106). The samples were tested by blinded observers using the modified and standard pH sticks. The sensitivities and specificities of the two pH sticks in identifying gastric and non-gastric aspirates were compared using the pH cut-off ≤5.5.
Results The sensitivities of a pH≤5.5 to correctly identify gastric samples were 66% (95% CI 56 to 75) and 68% (95% CI 57 to 77) for the modified and the standard pH, respectively. The specificities were 81% (95% CI 76 to 85) and 79% (95% CI 74 to 84). There were no significant differences in the distribution of the discordant results
between the paired gastric and non-gastric samples for both the modified and standard pH sticks at pH≤5.5 (both McNemar’s tests, p≥0.05).
Conclusions There were no significant differences between the paired modified and standard pH tests for the gastric samples. Due to the limited accuracy of pH sticks, further research is required to identify accurate and cost-effective bedside methods to confirm the correct placement of nasogastric tubes.

  • Type:

    Article

  • Date:

    31 July 2018

  • Publication Status:

    Published

  • Publisher

    BMJ

  • DOI:

    10.1136/bmjgast-2018-000218

  • Cross Ref:

    10.1136/bmjgast-2018-000218

  • Library of Congress:

    RT Nursing

  • Dewey Decimal Classification:

    610.73 Nursing

  • Funders:

    The University of Edinburgh

Citation

Rowat, A. M., Graham, C., & Dennis, M. (2018). Diagnostic accuracy of a pH stick, modified to detect gastric lipase, to confirm the correct placement of nasogastric tubes. BMJ Open Gastroenterology, 5(1), https://doi.org/10.1136/bmjgast-2018-000218

Authors

Keywords

nasogastric feeding tube, false negative pH tests, modified pH stick, gastric lipase,

Monthly Views:

Available Documents
  • pdf

    Diagnostic accuracy of a pH stick, modified to detect gastric lipase, to confirm the correct placement of nasogastric tubes

    643KB

    © Author(s) (or their employer(s)) 2018. Re-use permitted under CC BY-NC. No commercial re-use. See rights and permissions. Published by BMJ.
    This is an open access article distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited, appropriate credit is given, any changes made indicated, and the use is non-commercial. See: http://creativecommons.org/licenses/by-nc/4.0/

  • Downloadable citations

    HTML BIB RTF